CCD vs CMOS SNR calculator

In a 1 s exposure at 10 photons per pixel, a back-illuminated sCMOS (ORCA-Fusion BT, 1.0 e read noise) reaches an SNR of 2.8 against 2.2 for a deep-cooled CCD (iKon-M 934, 2.9 e). Past about 7 s the CCD's near-zero dark current wins: at 60 s it holds 2.2 while the air-cooled sCMOS falls to 1.1.

Winner Hamamatsu ORCA-Flash4.0 V3 C13440-20CU, by 8.054 dB.

A: Hamamatsu ORCA-Flash4.0 V3 C13440-20CU

Stacked SNR: 8.914 / 19 dB

Dynamic range (dB): 85.46

Dynamic range (stops): 14.19

Read-noise-limited signal (electrons): 2.56

B: Kodak (now onsemi) KAI-04050 monochrome ABA

Stacked SNR: 3.527 / 10.95 dB

Dynamic range (dB): 64.44

Dynamic range (stops): 10.7

Read-noise-limited signal (electrons): 144

Crossover (photons/pixel): No crossover in the plotted range

0.010.63139.825101.58e+51.00e+74.1e-40.0211.1603200Photons per pixel per exposureStacked SNR
A (solid): Hamamatsu ORCA-Flash4.0 V3 C13440-20CU; B (dashed): Kodak (now onsemi) KAI-04050 monochrome ABA; Ideal shot noise (grey)
Presets and curve values
Photons per pixel per exposureABIdeal shot noise
0.010.0050584.069e-40.1
0.0130.0065725.289e-40.114
0.01690.0085386.876e-40.13
0.021970.011098.938e-40.1482
0.028560.01440.0011620.169
0.037120.01870.001510.1927
0.048260.024260.0019630.2197
0.062730.031470.0025520.2505
0.081540.040790.0033180.2856
0.1060.052830.0043120.3256
0.13780.068350.0056060.3712
0.17910.08830.0072860.4232
0.23290.11390.0094710.4826
0.30270.14660.012310.5502
0.39350.18810.0160.6273
0.51150.24060.02080.7152
0.66490.30650.027030.8154
0.86440.38850.035120.9297
1.1240.48960.045641.06
1.4610.6130.059291.209
1.8990.76180.077021.378
2.4680.93920.11.571
3.2091.1480.12991.791
4.1711.3920.16862.042
5.4221.6730.21872.329
7.0491.9940.28352.655
9.1632.360.36733.027
11.912.7750.47543.451
15.483.2440.61453.935
20.133.7740.7934.486
26.174.3731.0215.115
34.015.0491.3125.832
44.215.8151.686.649
57.486.6822.1447.581
74.727.6652.7228.644
97.138.7813.4389.855
126.310.054.31411.24
164.111.495.37612.81
213.413.136.64614.61
277.414.998.14816.65
360.517.129.90518.99
468.719.5411.9421.65
609.322.2914.2724.68
79225.4316.9328.14
103029.0119.9532.09
133833.0923.3736.58
174037.7427.2341.71
226243.0331.5947.56
294049.0736.5154.22
382255.9642.0861.82
496863.8148.3970.49
645872.7555.5480.36
839682.9663.6691.63
1091094.5972.87104.5
14190107.883.34119.1
1844012395.25135.8
23970140.2108.8154.8
31170159.9124.2176.5
40510182.3141.8201.3
52660207.8161.8229.5
68460236.9184.6261.7
89000270.1210.6298.3
1.157e+5308240.2340.1
1.504e+5351.2273.9387.8
1.955e+5400.4312.4442.1
2.541e+5456.5356.3504.1
3.304e+5520.5406.2574.8
4.294e+5593.4463.2655.3
5.583e+5676.6528.2747.2
7.257e+5771.4602.2851.9
9.434e+5879.5686.7971.3
1.226e+610037831107
1.594e+61143892.71263
2.072e+6130410181440
2.694e+6148611611641
3.502e+6169513231871
4.552e+6193215092134
5.918e+6220317202433
7.693e+6251219612774
1.000e+7286422363162
A

Empty fields: not stated by the manufacturer. Fill in QE to calculate; empty dark current is treated as zero.

B

Empty fields: not stated by the manufacturer. Fill in QE to calculate; empty dark current is treated as zero.

Advanced
Join the Pro waiting list

What Pro produces

Batch runs over a table of inputs, a PDF report with inputs, method and citations, CSV export and saved cases. Join the waiting list and we will write once, when it opens.

Use this in your AI assistant

This calculator is available to Claude, ChatGPT and other assistants through ScienceShot’s free connector.

https://scienceshot.com/mcp

Connector tool: image_sensor_snr

Example: CCD vs CMOS SNR calculator

About the connector

Presets and curve values

SensorSource documentReadout mode
Andor iKon-M 934 BV (DU934P-BV) / 0.05 MHz; -80 C; QE up to 95%iKon-M 934 Series 5 MHz Imaging CCD 1 Megapixel, -100C (specification sheet)0.05 MHz; -80 C; QE up to 95%
Andor iXon Ultra 888 (#BV option) / EM amplifier 30 MHz; -80 C; QE >95%iXon Ultra EMCCD specification sheet (footer code iXonUltraSS 0425 R1)EM amplifier 30 MHz; -80 C; QE >95%
Hamamatsu ORCA-Flash4.0 V3 C13440-20CU / Standard scan; -10 C; QE at 560 nmTechnical note: ORCA-Flash4.0 V3 Digital CMOS camera C13440-20CUStandard scan; -10 C; QE at 560 nm
Hamamatsu ORCA-Fusion BT C15440-20UP / Standard scan; -8 C forced air; QE at 550 nmORCA-Fusion BT Digital CMOS camera C15440-20UP (Cat. No. SCAS0143E03)Standard scan; -8 C forced air; QE at 550 nm
Teledyne FLIR Blackfly S BFS-U3-51S5M (Sony IMX250) / EMVA 1288; 12-bit ADC; QE at 525 nmEMVA 1288 Imaging Performance, BFS-U3-51S5EMVA 1288; 12-bit ADC; QE at 525 nm
Kodak (now onsemi) KAI-04050 monochrome ABA / 40 MHz; photodiode dark current at 40 C; QE at 500 nmKODAK KAI-04050 Image Sensor, Product Summary40 MHz; photodiode dark current at 40 C; QE at 500 nm

Cite this

ScienceShot. CCD vs CMOS SNR calculator. https://scienceshot.com/ccd-vs-cmos-snr-calculator.

More formats

ScienceShot. (2026). CCD vs CMOS SNR calculator [Online calculator]. https://scienceshot.com/ccd-vs-cmos-snr-calculator

@misc{scienceshot-ccd-vs-cmos-snr-2026,
  author = {{ScienceShot}},
  title  = {{CCD vs CMOS SNR calculator}},
  year   = {2026},
  note   = {Online calculator},
  url    = {https://scienceshot.com/ccd-vs-cmos-snr-calculator}
}

Sources: EMVA 1288 Release 4.0 Linear (16 June 2021); An Overview of Photon Counting in EMCCD Cameras (Andor / Oxford Instruments, published 10 Feb 2021, updated 14 Feb 2022); Lucky imaging: beyond binary stars (arXiv 1404.5907); Hamamatsu ORCA-Flash4.0 V3 technical note, section 3-1 readout noise.